OpenWorldLab

Loading interactive view…

Diffraction

Resolving the sodium doublet

The yellow of a sodium street lamp is two colours: the D lines at 589.0 and 589.6 nm, a tenth of a percent apart. A prism cannot pull them apart; a diffraction grating can. Telling them apart is a standard test of a spectrometer.

What you are looking at

Parallel (collimated) sodium light falls on a grating ruled with 1200 lines per millimetre. Its first order leaves at 45°, where a lens of 500 mm focal length brings each line to a point on the screen.

The screen shows two bright spots about half a millimetre apart, one for each line, one twice as bright as the other: D₂ at 589.0 nm is the stronger. (With a slit in front of the lamp they would be lines.)

How the bench is set up

Sodium lamp behind a slit and collimator (0.3 mrad of spread, a 2 mm beam), a blazed grating of 1200 lines/mm, a 500 mm lens 150 mm from it, and a 4 mm screen at the lens’s focus.

The theory

A grating of line spacing dd sends light of wavelength λ\lambda into the angles where

dsin⁡θm=mλ.d\sin\theta_m = m\lambda.

Differentiating, two wavelengths Δλ\Delta\lambda apart leave Δθ=m Δλ/(dcos⁡θ)\Delta\theta = m\,\Delta\lambda / (d\cos\theta) apart, and a lens of focal length ff turns that angle into a distance on the screen:

Δx=f Δθ=f m Δλdcos⁡θ.\Delta x = f\,\Delta\theta = \frac{f\,m\,\Delta\lambda}{d\cos\theta}.

Here d=833d = 833 nm and θ=45∘\theta = 45^\circ, so 0.6 nm of separation becomes 1.0 mrad and Δx≈0.51\Delta x \approx 0.51 mm. Whether the two lines show as two depends on the resolving power R=λ/Δλ=mNR = \lambda/\Delta\lambda = mN, with NN the number of lines the beam covers: about 4800 here, against the 990 the doublet needs.

What to measure

Line separation f·Δλ/(d cos θ); resolving power mN

  1. Open the screen’s reading and measure the distance between the two spots.
  2. Compare with f Δλ/(dcos⁡θ)f\,\Delta\lambda/(d\cos\theta) = 0.51 mm, and check that the brighter line is the one at the shorter wavelength.

What the model leaves out

  • The slit and collimator are folded into the lamp: its light leaves with 0.3 mrad of spread, which sets how sharp each line is at the focus.
  • The grating is blazed and sends all the light into the first order.

Things to try

  • Raise the lamp’s divergence towards 2 mrad: each line spreads until the two merge into one. A wide slit costs resolution.
  • Change the grating to 600 lines/mm: the order comes out at 21°, and the lines move closer together.
  • Swap the gas for mercury and look for its yellow pair at 577 and 579 nm.

References

  • E. Hecht (2017), Optics, 5th edition, Pearson.

More experiments, and how the lab works, are in the laser lab’s overview.